Product Code | |
Personal LAB | 0-2875 |
Sample Station | |
Tubes | Up to 96 positions for tubes 12-13.5 mm diameter, 55 –75 mm height |
Label | UPC Version A and E (EAN 8 and 13), Code 39, Interleaved 2 and 5, Code 93, Code-bar, Discrete 2 or 5, Code 128, Code 39 FULL ASCII. The scanner will discriminate between the symbols, except Code 39 and Code 39 FULL ASCII. |
Loading Capacity | 96 samples |
Sample Identification |
Positive Sample Identification (PSID) by built-in barcode reader and dedicated sample loading facility |
Sample Loading | Sample strip loading with bar code reading |
Incubation Station | |
Incubation Area | Manages 2 microplates (with independent temperature control) |
Temperature | At least 5°C higher than the room temperature or 30°C ÷ 45°C |
Stability | ± 1°C |
Incubation Time | Programmable |
Warm-Up | 15 minutes |
Pipetting Station | |
Working Area | Manages 2 microplates independently |
XYZ Arm | 2 independent channels using disposable tips for sample, standards, controls and reagents |
Sample tip type | Adaltis disposable tips |
Tip size | 200 or 1000 μL |
Sample pipetting volume | 10-200 μL |
Reagent pipetting volume | 10-1000 μL |
Syringe Dilutors | Two dilutors with 1 ml and 5 ml precision micro-pumps |
Syringe 1 mL resolution | 4000 steps on max. stroke |
Syringes 5 mL resolution | 2000 steps on max. stroke |
Replicate Controls / Standards |
Programmable from 1 to 4 |
Replicate Samples | Programmable from 1 to 8 |
Precision for Serum Samples |
<5% with 10μL (CV), <3.5% with 25μL (CV) <2.0% with 50μL (CV), <2.0% with 100μL (CV) |
Precision for Reagents |
<3.5% with 50μL (CV), <2.5% with 100μL (CV), <2.5% with 200μL (CV) |
Serum Dispensing Time | <15 min. for 88 samples (volume 100μL) |
Reagent Dispensing Time | <3 min. for 96 wells (volume 100μL) |
Carryover | Disposable tips: none |
Washing Station |
|
Washing Volumes | 100 to 2000 μL |
Wash Cycles | 1 to 9 |
Soak time | 1 second to 3 minutes |
Pressure | Low or normal |
Photometric Reading | |
Working Area | Allows for 2 reading stations |
Reading Range | 0 to 3.0 OD |
Reading Channels | 16 (2X8) |
Reading methods | Single, double or double beam with over range filter vertical reading |
Reading Filters | 8 interference filters |
Standard Filter Range | 405-450-492-550-620 nm, (other filters optional) |
Linearity | 1% (0:2 OD) |
Precision | 1% (0:2 OD) |
Resolution | 0.001 OD |