IVD SYSTEMS & AUTOMATED CLINICAL DIAGNOSTIC ANALYSERS
NEXgen banner.jpg

Washing Station

Fully Automated - 7 Microplates - Open System

Washing Station

Manifold configuration

1 (8-way wash head)

Programmable volume

-50 - 2000 μL

Residual wash volume

< 3 μL per well in a flat bottom plate

Wash containers 6 tanks at 2 liters

6 tanks at 2.0 L 5 washing buffer + 1 distilled water or washing buffer

Buffer level alarm

Yes (with continuous level sensing)

Waste container

8 L with continuous level sensing

Dispense precision

≤5% CV (with 300 pL H2O in 96 well)

Liquid level sensing

Gravimetric for all tanks and waste container