Pipetting Station

Fully Automated - 7 Microplates - Open System

Pipetting Station

Number of pipettes

2 independent channels using disposable tips for sample and reagents

Sample tip type

Standard disposable tips

Sample tip size

200 -1000 μL

Sample pipetting volume

1 - 200 μL (with large tips up to 1000 μL)

Estimated transfer time

< 9 seconds per sample

Time to dispense

< 14 minutes

(for 96 samples with volume of 100 μL)

Precision

Reagents: ≤ 3% CV (10 Replicates) for any operating volume above 25 μL

Samples: ≤ 3% CV (10 replicates) for any operating volume above 10 μL

Accuracy

Reagents: ± 3% CV (10 replicates)
for any operating volume above 25 μL Samples:
± 3% CV (10 replicates) for any operating volume above 10 μL

Adding reagent
to full 96 plate

From 3 to 7 minutes depending on reagent volume (50 to 200 μL)